System Specifications

Perfusion Pumps

Voltage and Language Options
Catalog No. VAC Hz Language
043 605 00 115 60 English
043 605 001 230 50 English UK
043 605 002 230 50 French
043 605 005 230 50 Spanish
043 605 006 17 50 English
043 605 007 17 60 English
043 605 008 230 50 Italian
043 605 009 230 60 English
043 605 010 115 60 Spanish
Line voltage requirement is ± 10% VAC. Language refers to face labels only. Pumpheads can be rotated 90° clockwise (043
605 7XX) or counterclockwise
(043 605 SXX) prior to shipping.

DISPLAY ACCURACY (RPM)

± 2% of reading or ± 1 count whichever
is greater (without load).

TUBING SIZE RANGE-SILICONE RUBBER

Minimum 3/16" ID x 1/16" wall| Maximum 5/8" ID x 1/8" wall Double 1/8" - 1/4" ID x 1/16" wall non-continuous use only.

TUBING SIZE RANGE-PVC

Minimum 3/16" ID x 1/16" wall| Maximum 5/8" ID x 1/8" wall Double 1/8" - 1/4" ID x 1/16" wall non-continuous use only.

POWER CONSUMPTION

LOAD Usage Maximum 280 Watts Normal 150 Watts

PHYSICAL CHARACTERISTICS

Length: 56 cm (22.1 in) Width: 18 cm (7.1 in) Height: 33 cm (13.0 in) Weight: 26 kg (57 Ibs)

LEAKAGE CURRENT

100/115 VAC 12mA rms maximum 230 VAC 20 mA rms maximum

OCCLUSION

Adjustment 0.018 mm (0.707 in) per detent

STANDARD ACCESSORIES

Tubing Inserts Catalog No. 3/16" to 3/8" ID with 043 205 001 1/16" to 3/32" wall Buckberg 4:1 Plegia 043 206 004

PERFUSION PUMP OPTIONS

Tubing Inserts Catalog No. Dbl. 1/4" x 3/32" 043 205 000 Dbl. 5/16" x 1/16" 043 205 70 Dbl. 1/4" x 1/16" 043 205 002 Buckberg 1:1 Plegia 043 206 001 Buckberg 2:1 Plegia 043 206 002 Buckberg 8:1 Plegia 043 206 008

MASTER/SLAVE PUMP CABLE

Catalog Number 043028000

Console Bases

Console Base Size Selection
Catalog No. Width (cm/in.) Weight (kg/lb) Size
043 625 03X 69/27 44/97 3
043 625 04X 86/34 48/105 4
043 625 05X 104/41 51/113 5
043 625 06X 122/48 55/121 6
043 625 07X 140/55 7

ELECTRICAL CONFIIGURATIONS

X VAC HZ Typical Location 1 115 50/60 North America 2 230 50/60 United Kingdom 3 230 50/60 European Continent 4 100 50/60 Japan Line voltage requirements is ±10% VAC

CIRCUIT BREAKERS

14 Amp 17/115 VAC 10 Amp 230 VAC

STANDARD ACCESSORIES

Two IV poles/masts and one crossbar mast One pushbar (can be located on either side) Four locking casters and 20 cable dips 5-10 system electrical connectors 2-3 isolated accessory eledrical connectors Drip trays in between pumps Two hand cranks 10 Velcro strips.

CONSOLE BASE OPTIONS

Venous Line Occluding Clamp - Catalog Number 043 213 104; short cable for console bases of three positions or less. Catalog Number 043 213 105; long cable for console bases of four positions or more. Maximum tubing size is 3/4" OD. There is no minimum tubing size.

Computerized Profusion Controller

Voltage and Language Options
Catalog No. VAC Hz Language
043 655 500 115 50 / 60 English
043 655 501 230 50 / 60 English
043 655 502 230 50 / 60 French
043 655 505 230 50 / 60 Spanish
043 655 506 17 50 / 60 English
043 655 508 230 50 / 60 Italian
043 655 510 115 50 / 60 Spanish
Line voltage requirement is ± 10% VAC.

TIMERS - 3 Channels

Range: 00:00:00 to 99:59:59 (hours:minutes:seconds) Acuracy: ± 1% of elapsed time.

POWER CONSUMPTION

80 watts maximum

CIRCUIT BREAKERS

1.0 Amp at 100/115 VAC; 0.6 AMP at 230 VAC

LEAKAGE CURRENT

100/115 VAC 20 mA rms maximum 230 VAC 40 mA rms maximum

Computerized Profusion Controller (cont.)

TEMPERATURE MONITORS 6 CHANNELS

Display range: 0-50. C Display resolution: 0.1. Accuracy: ±1.0.C w/probe Transducer compatibility: YSI-400 and YSI-500 with calibration Pressure Monitor - 2 Channels Pressure Alarm range: -200 to +400 mmHg Pressure limited range -210 to +450 mmHg Accuracy ±8 mmHg w/transducer

AIR EMBOLI PROTECTION SYSTEM (AEPS)

AEPS Open Reservoir Level Detection Sensitivity: Within diameter of transducer (1 cm/0.4in.)~ 0 40 Hct. blood temerature
10-40'C

AEPS BUBBLE DETECTION SENSITIVITY

> 0.7 ml at flow rates < 6 5 Ipm with 3/8" x 3/32" tubing (PVC only) > 0.3 ml at flow rates <3.0 |pm with 1/4" x 3/32" tubing (PVC only) Response time for pump control: <1 second Arterial line tubing compatibility: 3/8" x 3/32" or 1/4" x 3/32" (PVC only)

CARDIOPLEGIA DELIVERY SYSTEM DISPLAY RANGE

Last volume delivered - 0 to 9999 ml Total volume delivered - 0 to 9999 ml Time since last delivery - 0 to 99:59:59 (hours: minutes:seconds) Accuracy: ±10% of desired volume

PHYSICAL CHARACTERISTICS
ELECTRONICS MODULE

Depth: 53 cm (21 in); Width 18 cm (7.1 in) Height" 31 cm (12 in); Weight 13.6 kg

DISPLAY MODULE

Depth 6 cm (2.4 in) w/out mount Depth: 22 cm (8.7 in) w/out fixed mount Width 34 cm (13.3 in); Height: 24 cm (9.3 in) Weight: 4 kg (9 Ibs) w/mount

STANDARD ACCESSORIES

Remote Control Cable - 10 ft (2 each) Catalog Number 043 255 018 Disposable Pressure Transducer (2 ea) Pressure Transducer Cable - 12 ft (2 ea) Air Bubble Sensor and Cable Catalog Number 600 005 100 Air Bubble Sensor Holder & Mount Catalog Number 600 064 000 AEPS 10 ft Control Cable - CPC to Pump Catalog Number 043 217 000 CPC Video Data Cable -10 ft Catalog Number 043 255 017 CPC Display Control Cable - 10 ft Catalog Number 043 218 000 Temperature Simulator Kit Catalog Number 043 252 100 CPC Display Arm - 10 cm (4 in) Catalog Number 043 255 016

Bio-Console & Blood Pumps

Bio-Console® 560 System

Indications for use: The Medtronic centrifugal blood pumping system is intended to pump blood through the extracorporeal bypass circuit for extracorporeal support for periods appropriate to cardiopulmonary bypass procedures up to six hours. Contraindications: The Medtronic blood pumping system is contraindicated as a cardiotomy suction device. This device used for any other purposes than for the indicated use is the responsibility of the user.

BPX 80 Bio-Pump® Plus Centrifugal Blood Pump

Indications for use: The Bio-pump is indicated for use only with the Medtronic Bio-Console pump speed controller to pump blood through the extracorporeal bypass circuit for periods appropriate to cardiopulmonary bypass procedures up to six hours.

Contraindications: The Medtronic blood pumping system is contraindicated as a cardiotomy suction device. This device used for any other purposes than for the indicated use is the responsibility of the user.

AffinityTM CP Centrifugal Blood Pump

Indications for use: The Affinity CP Centrifugal Blood Pump is used to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal support systems (for periods up to 6 hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants). The Affinity CP Centrifugal Blood Pump is driven by the External Drive Motor or the Emergency Handcrank.

Contraindications: The Affinity CP Centrifugal Blood Pump is contraindicated for use as a cardiotomy suction device. This device, used for any other purposes than the intended use, is the responsibility of the user.

Caution: Federal law (USA) restricts these devices to sale by or on the order of a physician. For a complete listing of indications, contraindications, precautions and warnings, please refer to the Instructions for Use which accompany each product.